DSpace Repository

Comparison of the automatd full blood count results with manual method

Show simple item record

dc.contributor.author Threeswaran, R.
dc.contributor.author Sooriyakumar, T.
dc.contributor.author Arasaratnam, V.
dc.date.accessioned 2021-07-13T08:42:30Z
dc.date.accessioned 2022-07-07T07:14:22Z
dc.date.available 2021-07-13T08:42:30Z
dc.date.available 2022-07-07T07:14:22Z
dc.date.issued 2012
dc.identifier.issn 2279-1922
dc.identifier.uri http://repo.lib.jfn.ac.lk/ujrr/handle/123456789/3484
dc.description.abstract This study was designed to determine the correlation between heamatological parameters measured by an Automated Haematology analyzer (MINDRAY BC: 5500) with the standard manual methods. In this study, One hundred and fourty Anticoagulated (tri-potassium ethylenediamine tetra-acetic acid) blood samples from patient having different blood disorders were randomly selected from Haematology Laboratory Teaching Hospital Jaffna, which were used. All the blood samples were tested automatically by automated haematology analyzer as well as manually for WBC, RBC, Platelet, Haemoglobin, PCV, differential count, MCV, MCH and MCHC. All the data were statistically analyzed in SPSS software. The results obtained with haematology analyzer for WBC count, Haemoglobin measurement, PCV, RBC count, Platelet count, Neutrophil count, Lymphocyte counts, Eosinophil count, Basophil count, Monocyte count, MCV, MCH and MCHC were correlated with manual methods (Pearson correlation respectively 0.920, 0.895, 0.865, 0.662, 0.699, 0.724, 0.70, 0.456, 0.347, 0.087, 0.308, 0.462 and 0.117). The results obtained with automated haematology analyzer for WBC count PCV and Hb well correlated with standard manual method than other haematological parameters. Further studies are required for proper quality control and calibration to optimizing the validity of full blood count results. en_US
dc.language.iso en en_US
dc.publisher University of Jaffna en_US
dc.subject Full blood count en_US
dc.subject Automated haematology analyzer en_US
dc.title Comparison of the automatd full blood count results with manual method en_US
dc.type Article en_US