Erroneous intracranial pressure measurements from simultaneous pressure monitoring and ventricular drainage catheters

Show simple item record

dc.contributor.author Birch, A.A
dc.contributor.author Eynon, C.A
dc.contributor.author Schley, D
dc.date.accessioned 2014-02-01T16:31:55Z
dc.date.accessioned 2022-06-28T06:46:01Z
dc.date.available 2014-02-01T16:31:55Z
dc.date.available 2022-06-28T06:46:01Z
dc.date.issued 2006-08
dc.identifier.issn 15416933
dc.identifier.uri http://repo.lib.jfn.ac.lk/ujrr/handle/123456789/178
dc.description.abstract The objective of this report is to highlight the potential for false pressure measurements from systems that combine intracranial pressure (ICP) measurement and ventricular drainage. If the ports of the drain become blocked to the extent that they present a high resistance to cerebrospinal fluid flow, then a significant pressure gradient between the inside and outside of the catheter may be established. Thus, any intracatheter transducer will faithfully record a pressure much lower than true ICP. This holds true for catheter-tip transducers when the transducer lies inside the catheter. In the absence of flow, however, pressures will equalize; therefore, accurate measurements may be taken if the drain is temporarily closed. We model this situation and provide simulations of expected measurements in such situations; these compare well to observed clinical readings. en_US
dc.language.iso en en_US
dc.publisher Humana Press Inc en_US
dc.subject Cerebrospinal fluid en_US
dc.subject Intracranial pressure en_US
dc.subject Ventricular drainage en_US
dc.title Erroneous intracranial pressure measurements from simultaneous pressure monitoring and ventricular drainage catheters en_US
dc.type Article en_US